# How EBSD Parameters Are Decided

This note explains *why* the acquisition parameters recommended for Electron
Backscatter Diffraction (EBSD) take the values they do — the physics and
statistics that fixed the settled optima recorded in standards such as
ASTM E2627 and ISO 13067, rather than simply pointing at those standards.

Every EBSD parameter is a point on an explicit trade-off curve balancing
**pattern signal/contrast** vs. **spatial resolution** vs. **angular precision**
vs. **acquisition speed** vs. **statistical representativeness**. The underlying
physics (backscatter yield, interaction volume, `λ ∝ 1/√V`, `SNR ∝ √N`, gnomonic
projection geometry) fixes the *shape* of each curve; sampling and confidence
statistics fix *how far along* it you need to go.

> Note on provenance: the reasoning below is derived from established EBSD
> physics and metrology. It was **not** produced via Edison Scientific — that
> path is currently blocked by an empty `EDISON_PLATFORM_API_KEY` in CI. Once a
> non-empty key is provided, the same query can be re-run through the
> `LITERATURE_HIGH` (paperqa3-high) job to produce a fully citation-grounded
> version, and the trajectory artifacts committed alongside this document.
> Verify the exact numeric clauses of ASTM E2627 / ISO 13067 against current
> published editions before using them for compliance.

## 1. Specimen tilt ≈ 70°

The backscatter coefficient η rises steeply with tilt, and — more importantly —
the *fraction* of backscattered electrons that retain coherent diffraction
information (the Kikuchi signal, versus diffuse background) peaks in roughly the
60–75° range. Tilting shortens the escape depth normal to the surface, so the
emerging electrons sample a thin near-surface layer where their exit
trajectories still carry channeling/diffraction contrast. Above ~70° the pattern
contrast keeps improving marginally, but the projected beam footprint elongates
(spatial resolution becomes strongly anisotropic), geometric
foreshortening/distortion grows, and shadowing worsens. **70° is the empirical
optimum** balancing Kikuchi-band contrast against spatial resolution and
geometric distortion — that is why the standards fixed on it, not the reverse.

## 2. Accelerating voltage (typ. 10–30 kV, 20 kV default)

Two competing physics chains set this:

- **Signal / pattern sharpness:** electron wavelength `λ ∝ 1/√V`, and Kikuchi
band width scales as `~2λ/d` (Bragg angle). Higher kV → shorter λ → thinner,
sharper bands and higher BSE yield → better statistics and, if the detector
can resolve the thinner bands, better angular precision.
- **Spatial resolution:** higher kV enlarges the interaction/excitation volume
(deeper, wider), degrading resolution and blurring measurements near
boundaries.

So 20 kV is a general-purpose compromise; you drop to ~10 kV for nanoscale
features, thin films, or beam-sensitive/low-Z materials (smaller interaction
volume), and push to 30 kV for weakly scattering light-element phases that need
more signal. Low-Z elements have low BSE yield → need higher kV/current; high-Z
phases scatter strongly → tolerate lower kV.

## 3. Beam / probe current

Pattern SNR `∝ √N` (electrons per pattern), so more current gives cleaner
patterns and shorter exposure (faster maps). The cost is probe size: for a given
column, higher current means a larger probe → worse spatial resolution (less
severe on FEG than thermionic sources). You raise current for speed/SNR on
coarse microstructures and lower it (small probe) to resolve fine grains and
boundaries.

## 4. Step size — the "≥ ~10 points per grain" rule

This comes from sampling statistics, not convention. If a grain is covered by N
measurement points, the fractional error in its reconstructed area/diameter from
pixel discretization falls roughly as `~1/√N`, and boundary/grain-count
reconstruction becomes unreliable when the smallest grains contain only a
handful of points. Choosing step ≤ (smallest grain of interest)/10 puts ≥ ~10
points across a grain, bringing pixelation error down to a few percent and
letting a grain-reconstruction algorithm (misorientation-threshold flood fill,
typ. 5°) distinguish real grains from noise. **ASTM E2627** codifies exactly this
logic (minimum points-per-grain and a misorientation threshold for defining a
grain) so that grain-size measurements are reproducible. The lower bound on step
size is physical: nothing is gained by stepping below the effective spatial
resolution (~20–50 nm for FEG at 20 kV) or below what intragranular-gradient
(GND) sensitivity requires.

## 5. Working distance, detector distance, and pattern centre

WD (~15–20 mm) is set so the tilted region of interest projects onto the phosphor
with correct geometry. Detector insertion distance trades **angular capture**
against **gnomonic magnification**: a closer detector captures a wider slice of
the diffraction sphere (more bands → more over-determined, more robust indexing)
but lower magnification per band. The **pattern centre** must be calibrated
because indexing solves orientation from band positions *relative to the PC* — a
PC error maps almost linearly into an orientation error, which is why
calibration (known-orientation standard, moving-screen, or iterative refinement)
is mandatory rather than optional.

## 6. Camera binning, exposure, gain

Detector pixel count sets the ceiling on angular resolution: band-position
precision improves with more pixels across the pattern, so orientation precision
does too — but at lower per-pixel SNR and slower readout. **Binning** (4×4, 8×8)
sums pixels for higher SNR and frame rate at the cost of angular resolution:
heavy binning for fast mapping, low/no binning for high-precision or HR-EBSD
(resolving <0.1° misorientations). Exposure/frame-averaging improves SNR as
`~√(frames)`, sharpening Hough peaks and indexing reliability; you set it to the
minimum that reliably meets your indexing-quality target (CI/MAD) at the desired
speed. Gain amplifies signal *and* noise, so it doesn't substitute for real
exposure.

## 7. Hough transform parameters

More detected bands over-determine the orientation fit → higher precision and
fewer misindexes, with diminishing returns and slower indexing (typ. 7–12
bands). Finer Hough (θ, ρ) binning improves band-position precision (→ better
angular resolution) but is noisier and slower; the butterfly-mask size is tuned
to the expected band width, which itself depends on kV and detector distance.
These are chosen to sit at the knee of the speed / indexing-success / precision
trade-off for the material.

## 8. Minimum grains / scan area for representative statistics

The counts in standards are confidence-interval arithmetic. The standard error
of the mean grain size `∝ σ/√(n_grains)`, so to bound the mean to ±X% at 95%
confidence you need `n ≥ (1.96·σ / (X·mean))²` grains — a few hundred to a few
thousand depending on size spread. Texture (ODF) representativeness is stronger
still (typically thousands of grains for a stable ODF, more for weak/random
textures). The required scan area then follows from
(grains needed) × (mean grain area), subject to the step-size constraint — which
is what ultimately sets map size and acquisition time.

## 9. CI / MAD / band-contrast thresholds for clean-up

These cutoffs were set empirically by correlating each metric against
independently confirmed correct indexing:

- **Confidence Index** (TSL) `= (V₁−V₂)/V_ideal` from the band-voting scheme;
**CI > 0.1** is the conventional "reliable" line because, across many
materials, >99% of points above it index correctly.
- **MAD** (mean angular deviation, Oxford) is the mean misfit between detected
and simulated bands; **MAD < ~1°** marks a good fit.
- **Band-contrast / image-quality** thresholds separate poorly diffracting
points (boundaries, deformed regions, unindexed phases).

Clean-up routines (neighbour-orientation correlation, grain dilation) use these
thresholds to drop/replace unreliable points — and standards require reporting
the clean-up because over-aggressive filtering biases grain-size and texture
results.

## Summary

The through-line is that every EBSD parameter sits on a physics- and
statistics-defined trade-off curve. The physics fixes the shape of the curve;
sampling and confidence statistics fix how far along it you must go. The
standards simply record the settled optima of those curves for common materials
— which is exactly the reasoning CALIBER aims to make explicit and
sample-specific when recommending baseline acquisition parameters.
Binary file added docs/images/convention-validation.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/images/euler-angles.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/images/eutectic-si-schematic.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/images/gnomonic-projection.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added example/area2-results/area2_full.ang.gz
Binary file not shown.
Binary file added example/area2-results/ci_hist.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file not shown.
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
36 changes: 36 additions & 0 deletions example/area2-results/dictionary-indexing/stats.json
Original file line number Diff line number Diff line change
@@ -0,0 +1,36 @@
{
"n_patterns": 4900,
"dictionary_size": 52607,
"dictionary_resolution_deg": 2.5,
"pc_bruker": [
0.531,
0.3204,
0.7338
],
"hough": {
"s_per_map": 10.6,
"ci_gt_0.1": 0.9927,
"median_ci": 0.672,
"median_fit_deg": 0.514
},
"di_al": {
"s_projection": 13.0,
"s_indexing": 48.4,
"s_refinement": 87.3,
"median_ncc_di": 0.221,
"median_ncc_refined": 0.285
},
"agreement_hough_vs_di": {
"median_misorientation_deg": 0.78,
"frac_within_2deg": 0.794,
"frac_within_5deg": 0.831,
"median_misori_where_ci_lt_0.1": 41.42,
"n_ci_lt_0.1": 36
},
"phase_test_al_vs_si": {
"median_ncc_al": 0.285,
"median_ncc_si": 0.209,
"frac_al_wins": 0.991,
"median_delta_ncc": 0.073
}
}